|
|
|
|
|

|
Protocol
Rapid quench operational procedures
Preliminary procedures
- Prepare buffers.
- Turn on the thermometer and the water bath for 25 °C
- Start mixing the enzyme/DNA and the dNTP/Mg2+ to let them sit on ice for 20 min.
- Turn on the controller. Plug in the adapter cord to the keyboard. Disregard the message "system error 5" if it appears. Type in "RUN," "return"
- When the main menu pulls up, type "7" for seeking home position of the syringe driver.
- Load Syringes.
- Turn syringe knobs to Load position.
- Add 1 Reaction Buffer without Mg2+ to left and right syringes.
- Load 500 mM EDTA (pH 8.0) to Quench syringe.
- Keep pushing in and out the plungers to get rid of the bubbles inside the syringes
- Change the syringe knobs to "fire" position.
- Type "2" for adjusting position of the syringe bar. Press "start" button on the controller to lower (+, default) or raise (-) the bar. Keep lowering the syringes till the buffers start coming off the exit loop.
- Type "ESC" to return to the main menu.
- Fill the enzyme/DNA and dNTP/Mg2+ mixes respectively into 1 mL disposable syringes. Plug the syringes into the designated sample load valves.
- Switch the port to the required loop. Repeat the "wash" procedure.
Collect time point
- On the main menu, type "1" for quench-flow run.
- "Enter the reaction time (sec)" type any value larger than 0.002s. "Return." Then the corresponding loop number will be displayed.
- Determine which reaction loop is required
- Move to new reaction loop.
- Wash the loops.
- Turn on vacuum
- Quench knob to off position. Left knob to off. Right knob to flush.
- Place H2O under right flush line.
- Attach vacuum to center line (sample exit loop)
- Let H2O flow for 10 sec.
- Move H2O from right line to left line
- Left knob to flush. Right knob to off.
- Let H2O flow for 10 sec.
- Place right line in MeOH
- Right knob to flush. Left knob to off.
- MeOH for 5 s
- Move MeOH from right ot left line.
- eft knob to flush. Right knob to off.
- MeOH for 5 s
- Right knob to flush
- Remove MeOH
- Let lops dry for 15 s.
- Turn Sample Values to load.
- Load DNA/enzyme into sample load loop
- Load dNTP/Mg2+ into opposite sample load loop.
- Turn Sample Valves to Fire.
- Check that all valves are in Fire position.
- Place microcentrifuge tube onto end of sample exit loop.
- Press "G" to run reaction.
- Remove microcentrifuge tube.
- Go to step 2.
- When done with all the samples, enter "0" for reaction time to return to the main menu.
- Then enter "7" for home position of the syringe driver
- Exception: Do NOT load the dATP/Mg2+ when quenching at 0s time point.
Wash up.
Wash syringes.
- Turn Syringe valves to Load.
- Press all 3 syringe-plungers to retrieve the 1X Reaction Buffer without Mg2+ and 500mM EDTA buffer. Recap and store away the buffers.
- Wash syringes with H2O three times.
- Fill syringes for a fourth time and switch Syringe knobs to Fire.
- Push H2O through each of the reaction loops.
Sample loading loops
- Remove dNTP/Mg2+ and enzyme/DNA syringes
- Switch Enzyme/DNA knob to load.
- Place waste tube under Sample /exit Loop.
- Wash with
- 10mL 0.2N NaOH
- 10mL of 0.3N H3PO4
- 20mL H2O
- 1mL of MeOH.
- Turn knob to Fire.
- Turn dNTP/Mg2+ knob to load.
- Wash with
- 10 mL H2O
- 1 mL MeOH
- Turn knob to Fire.
Wash Reaction Loops
- Change knob positions and turn on the vacuum to expel the ddH2O.
- Lastly flush all the loops by dipping the flush loops into ddH2O for 10s and MeOH for 5s. Let dry.
- Turn off the controller and unplug the keyboard.
- Turn off the thermometer and water bath.
- Disconnect the water hose to drain the quench-chamber.
|
|
Back
|
|
|
|
College of Medicine
|
Medical Center
| Children's Hospital
Privacy and Legal Notices
|
|

|